The Memmert Pass-through oven UF450TS heating oven with fully insulated stainless steel doors on both sides saves time during loading and reduces the risk of contamination, especially when handling sensitive load between grey and clean rooms. This chamber comes standard with programmable AtmoCONTROL software and 2 stainless steel grid shelves.
Description
Temperature:
•Working temperature range at least 10 °C above ambient temperature to +250 °C
•Setting resolution temperature up to 99.9 °C: 0.1 / from 100 °C: 0.5
•Setting temperature range +20 to +250°C
•Resolution of display for actual values0.1°C
•Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Control Technology:
•ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays.
•Language setting German, English, Spanish, French, Polish, Czech, Hungarian
•Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
•Function HeatBALANCE adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
•Function SetpointWAIT the process time does not start until the set temperature is reached
•Calibration three freely selectable temperature values
Ventilation:
•Fan forced air circulation by quite air turbine, adjustable in 10 % steps for each segment individually
•Fresh air Admixture of pre-heated fresh air by electronically adjustable air flap
•Vent connection with restrictor flap
Communication:
•Documentation program stored in case of power failure
•Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programs via Ethernet interface or USB port
Safety
•Temperature control independently working, digitally adjustable electronic micro-processor overtemperature monitor TWW, protection class 3.1 (max-value for overtemperature, min-value for undertemperature)
•Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
•AutoSAFETY additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
•Autodiagnostic system for fault analysis
•Alarm visual and acoustic