Memmert Humidity Chamber – HCP provides excellent control, making it easy to achieve uniform atmosphere and temperature distribution in simulating environmental condistions. This 156L (5.4 cu ft) model is the 115 volt, with working humidity ranges of 20 to 95% rh, within temperatures of at least 6°C over ambient up to 90°C with humidity control. Includes (2) Stainless Steel Perforated Shelf, Atmocontrol Software + Manual, Operation manual, Calibration Certificate, Power Cord, Water tank and connections.

Description

Temperature:
• +7 above ambient temperature up to +90 °C
• Setting temperature range: +18 to +90 °C
• Setting accuracy temperature 0.1 °C
• Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity:
• Setting range humidity 20 to 95 % rh and rh-Off
• Setting accuracy humidity 0.5 % rh
Control Technology:
• TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays.
• Language setting German, English, Spanish, French, Polish, Czech, Hungarian
• Adjustable parameters temperature (Celsius or Fahrenheit), relative humidity, timer
• Timer digital backwards counter with target time setting, adjustable from 1 minute to 99 days
• Function SetpointWAIT the process time does not start until the set temperature is reached
• Calibration three freely selectable values each, temperature and humidity
Communication:
• Communication interface Ethernet LAN, USB
• Documentation program stored in case of power failure
• Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programs via Ethernet interface or USB port
Safety:
• AutoSAFETY additionally integrated over- and under temperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or under temperature, heating function is switched off in case of overtemperature, cooling function in case of under temperature
• Autodiagnostic system integral fault diagnostics for temperature and humidity control
• Alarm visual and acoustic
Heating Concept:
• large-area multi-function heating system on four sides with additional door and back heating to avoid condensation

Additional information

Voltage Configuration

115V, 230V