$0.00
Memmert Peltier Cooled Incubator (IPPeco PLUS) – this programmable 256 liter model is the 230 volt or 115 volt, Peltier-technology seamlessly heats and cools, with TwinDISPLAY digital controller. IPPeco PLUS temperature range 0°C to 70°C. Each individual appliance is tested according to the requirements of DIN 12 880: 2007-05. Includes (2) Stainless Steel Grid Shelves, Atmocontrol Software, Atmocontrol Manual, Operation manual, Calibration Certificate and Power Cord – energy efficient, precise.
Description
Temperature:
•Setting temperature range 0 to +70 °C
•Working temperature range without light: from 0 (at least 20 below ambient temperature) to +70°C
•Setting resolution temperature 0.1 °C
•Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Control Technology:
•ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays.
•Language setting German, English, Spanish, French, Polish, Czech, Hungarian
•Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
•Function SetpointWAIT the process time does not start until the set temperature is reached
•Calibration three freely selectable temperature values adjustable parameters temperature (Celsius or Fahrenheit), timer
Ventilation:
•Convection forced ventilation by Peltier fan
Communication:
•Documentation program stored in case of power failure
•Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programs via Ethernet interface or USB port
Safety:
•Temperature control over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
•AutoSAFETY additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
•Autodiagnostic system for fault analysis
•Alarm visual and acoustic
Heating Concept:
•Peltier energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)
Additional information
Voltage Configuration | 115V, 230V |
---|